| |
TM5-4120-376-14
b. Brazing.
(1) Heat evenly to recommended temperature. Keep the torch moving constantly in a
figure-eight motion.
(2) Apply silver brazing alloy to the heated parts.
Do not heat (melt) the silver brazing
alloy with the torch.
(3) Cool the joint.
(4) Clean the joint, using warm water and a brush.
(5) Be sure all flux has been removed.
5-11. LEAK CHECKING.
CAUTION
Rapid discharge will cause oil to be blown out of the compressor.
NOTE
The electronic leak detector is sensitive to the presence of refrigerant gas
(Table D-1, Item 11) in the atmosphere. When refrigerant gas is present in the
atmosphere of the work area, false indications can result. Use in a well
ventilated but draft-free area.
a. Connect a manifold gauge to the suction service valve on the unit and a cylinder of
refrigerant (Table D-1, Item 11). Open both service valves and the cylinder shutoff valve. Let
refrigerant flow into the system until the pressure gauge indicates 50 psig (3.5 kg/cm2). Close
cylinder shutoff valve and suction service valve and disconnect the refrigerant cylinder.
b. Connect a cylinder of dry nitrogen (Table D-1, Item 8) to the suction service valve. Open
the nitrogen cylinder shutoff valve and the suction service valve and pressurize the system to
350 psig (22 kg/cm2). Close both valves and test for leaks using an electronic leak detector or
the soap bubble method as described below:
(1) Electronic Leak Detector. Turn the electronic unit on and slowly pass the probe
around all points of the refrigeration system at which a leak could exist. Depending upon the
type of detector used, leak will be indicated by an audible signal, a light, or by meter
deflection.
(2) Soap Solution. Brush soap solution on all possible points of leakage and watch for
bubbles. Follow a definite sequence to avoid missing any joints that should be tested. Wipe the
solution from all joints and mark any point at which a leak is found.
(3) Discharge the system after leak checking by connecting a hose to the suction service
valve and cracking the valve open slightly to slowly discharge the gas. If leaks were detected,
repair them and retest in accordance with paragraph 5-11. If the system is tight, double
evacuate and charge the system in accordance with paragraphs 5-12 and 5-13.
5-13
|