|
|
TM 5-4120-369-14
SERVICE VALVES - Continued.
5-17.
Installation.
b.
(1) Refer to figure 5-1A and mount service valve.
(2) Attach refrigerant line flare nut to service valve.
Caution should be exercised with fluorocarbon
refrigerant gas as they can displace oxygen and
thereby cause suffocation.
Personnel with a history or other evidence of cardiac
rhythm abnormalities should be made aware of the
potential for aggravation of existing cardiac rhythm
abnormalities or the induction of arrhythmias, as a
result of exposure to fluorocarbon refrigerant gases.
Such individuals should be evaluated by local medical
authorities before working in environments where
potential refrigerant exposure may occur.
(3) Refer to figure 6-3 and discharge tine refrigeration system.
(4) Refer to figure 5-1A and disconnect the service valve flare nuts.
(5) Refer to paragraph 6-3 and leak test the system.
(6) Refer to figure 6-3 and purge the refrigeration system.
(7) Refer to figure 6-3 and charge the refrigeration system.
(8) Refer to paragraph 4-18 and install top covers.
HIGH PRESSURE CUT OUT (HPCO) AND LOW PRESSURE CUTOUT (LPCO)
SWITCHES.
Removal .
a.
(1) Refer to para 4-18 and remove the top covers.
Caution should be exercised with fluorocarbon refrigerant
gas (freons) as they can displace oxygen and thereby
cause suffocation.
personnel with a history or other evidence of cardiac
rhythm abnormalities should be made aware of the potential
for aggravation of existing cardiac rhythm abnormalities
or the induction of arrhythmias, as a result of exposure to
fluorocarbon refrigerant gases (freons) . Such individuals
should be evaluated by local medical authorities before
working in environments where potential freon exposure
may occur.
(2) Refer to paragraph 6-3 and discharge the refrigerant system.
(3) Refer to paragraph 5-17 and disconnect the service valve flare nuts.
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |