|
|
TM 9-4120-430-14
0048 00
NOTE
In the event the compressor was replaced as a result of burn-out, check that
compressor burn-out procedures were followed. See WP 0059 00.
SERVICE
EVACUATION
1.
Connect the hose from the low pressure service valve to the compound gauge side of the charging manifold. The
hose from the high pressure service valve shall be connected to the high pressure gauge side of the charging
manifold. See figure 1.
2.
Open both service valves.
3.
Attach center hose assembly charging manifold to vacuum pump.
4.
Start vacuum pump.
5.
Open charging manifold valves.
6.
Run the vacuum pump until approximately a 300 micron vacuum is reached.
NOTE
Inability to reach 300 microns may indicate either a leak or a problem with the
pump.
7.
Close manifold valves and check compound gauge. Record reading. Let unit sit for one hour. Observe compound
gauge reading.
8.
If the system holds the vacuum without change of pressure, proceed to step 10.
9.
If the vacuum cannot be held for one hour, one of the following reasons may account for the problem.
a.
Presence of water vapor in the system. Continued pumping will correct this condition.
b.
Leak in the refrigeration system. Break the vacuum with dry nitrogen and retest for leaks. See WP 0047 00.
c.
Internal leakage of vacuum pump. Test the pump by connecting a vacuum gauge directly to the vacuum pump
intake and continue to pump. If pump still fails to reach approximately 300 microns, the pump is faulty.
10. Close charging manifold valves.
11. Stop vacuum pump.
12. Disconnect pump from center hose connection.
13. Charge system with refrigerant-22. See WP 0049 00.
0048 00-2
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |